Mainstream Anesthesia Multi-Gas Module, simultaneously measure the concentration waveform and data of carbon dioxide, nitrous oxide and anesthetic gas, as well as respiratory rate. The types of anesthetic gas measured include halothane, enflurane, isoflurane, sevoflurane and desflurane. The user can manually select the type of anesthetic gas.
Equipped With Oxygen Battery. Reliable multi gas OEM solution, compatible with a variety of protocols, so that your products can be quickly upgraded and assembled.
Can simultaneously measure the concentration waveform and data of carbon dioxide, nitrous oxide and anesthetic gas, as well as respiratory rate. The types of anesthetic gas measured include halothane, enflurane, isoflurane, sevoflurane and desflurane. The user can manually select the type of anesthetic gas.
The host does not need internal circuit, just a serial port. The data can be measured within 4 seconds after power on. It is suitable for adults, children and newborns, with excellent cost performance and low cost.
There is an atmospheric pressure sensor inside, which is suitable for various altitude ranges. High-precision. Careful selection of infrared devices, hardware design and software algorithm, to achieve high signal-to-noise ratio, to ensure the measurement accuracy for temperature range of - 5-50 °C and in the range of 0 - 20%.
Working condition | 5°C to 50°C, 10 to 90% RH, noncondensing |
Circumstance pressure | 55-115kpa |
Power consumption | <180 mA |
Interface | RS-232/Standard E-8-1, 38.4 kBaud,10 ms data interval (KM) or customized communication protocol |
EtCO2, FiCO2 measurement | range:0-150 mmHg, 0-20 kPa, 0-19.7% Accuracy:±2mmHg for CO2 range 0 - 40mmHg ±5% for CO2 range from 41 - 70mmHg ±8% for CO2 range from 71 - 100mmHg ±10% Over 100mmHg |
FiN2O, EtN2O measurement | Range: 0-100% Accuracy: ± (2 vol% + 2% of reading |
FiHAL/EtHAL, FiENF/EtENF, FiISO/EtISO measurement | Range: 0-12% Accuracy: 0-8%: ± (0.15 vol% + 5% of reading) 8-12%: unspecified |
FiSEV, EtSEV measurement | Range: 0-15% Accuracy: 0-10%: ± (0.15 vol% + 5% of reading) 10-15%: unspecified |
FiDES, EtDES measurement | Range:0-25% Accuracy: 0-18%: ± (0.15 vol% + 5% of reading) 18-25%: unspecified |
Gas compensation | O2 (HOST supplied), or automatically compensation if O2 equipped |
Barometric Pressure | Having barometric pressure sensor and automatically compensation |
Warm-up time | 10 seconds reach designed deviation 97%, 20 seconds reach designed deviation. |
Rise times (t10- 90 %) | CO2:<70ms N2O and AGENT:<300ms |
Calibration | No need to calibrate |
Zero | Implement when replacing adaptor |
Equipped with Oxygen battery, reliable multi gas OEM solution, compatible with a variety of protocols, so that your products can be quickly upgraded and assembled.
Mainstream mode, can simultaneously measure the concentration waveform and data of carbon dioxide, nitrous oxide and anesthetic gas, as well as respiratory rate. The types of anesthetic gas measured include halothane, enflurane, isoflurane, sevoflurane and desflurane. The user can manually select the type of anesthetic gas.
Plug and play, the host does not need internal circuit, just a serial port. The data can be measured within 4 seconds after power on. It is suitable for adults, children and newborns, with excellent cost performance and low cost.
Automatic pressure compensation. There is an atmospheric pressure sensor inside, which is suitable for various altitude ranges. High-precision. Careful selection of infrared devices, hardware design and software algorithm, to achieve high signal-to-noise ratio, to ensure the measurement accuracy for temperature range of - 5-50 °C and in the range of 0 - 20%.